Calibration Certificate

Northgate Calibration Ltd
UKAS 0417
NCL-2026-00001
Revision 1 · Issued 2026-09-09
Calibrated by: J. Moreno   Reviewed by: J. Moreno   Environment: Relative humidity 44 %RH · Temperature 20.3 °C
Instrument: Digital caliper 0–150 mm
Asset Tag: CAL-0091
Manufacturer: Mitutoyo
Model: 500-196-30
Serial: A2318804
Calibration Date: 2026-09-09
Decision Rule: guardband
Worst-case TUR: 1.73
Result: CONDITIONAL
Condition as received: in tolerance.

Calibration Results

PointNominalTolerance / limitsAs-Found (mm)Error (AF) (mm)As-Left (mm)Expanded UTURStatus (as-left)
10 mm 10 mm± 0.02 mm
9.98 – 10.02 mm
10.003 0.003 10.003 ± 0.012 mm (k=2)1.73:1 PASS
50 mm 50 mm± 0.02 mm
49.98 – 50.02 mm
50.012 0.012 50.012 ± 0.012 mm (k=2)1.73:1 CONDITIONAL
specification met (within tolerance) but inside the guard band, so NOT accepted under the guard-banded rule; probability of non-conformance if treated as accepted: 8.3% (normal, u = U/k)
100 mm 100 mm± 0.02 mm
99.98 – 100.02 mm
100.019 0.019 100.019 ± 0.012 mm (k=2)1.73:1 CONDITIONAL
specification met (within tolerance) but inside the guard band, so NOT accepted under the guard-banded rule; probability of non-conformance if treated as accepted: 43.1% (normal, u = U/k)
150 mm 150 mm± 0.03 mm
149.97 – 150.03 mm
150.011 0.011 150.011 ± 0.012 mm (k=2)2.59:1 PASS

Decision rule (ISO/IEC 17025 §7.8.6)

Guard-banded acceptance (ILAC-G8:2019 §4.2.2 / ANSI-NCSL Z540.3 method 6): a result is reported PASS only when it lies inside the tolerance by at least the guard band w = U (U = expanded uncertainty, k as stated per point); a result within tolerance but inside the guard band is reported CONDITIONAL with its probability of false acceptance (PFA, JCGM 106); a result outside tolerance is reported FAIL. Probabilities assume a normal distribution with standard uncertainty u = U/k about the measured value (JCGM 106). Rule agreed with the customer before calibration.

Calibration conditions

ProcedureNCL-DIM-004 · Calibration of digital calipers to 150 mm · rev 1
EnvironmentRelative humidity 44 %RH · Temperature 20.3 °C
Locationlaboratory premises
Customernot recorded
Item receivednot recorded (calibrated without an inbound order)

Reference standards (traceability)

DescriptionCertificateTraceable toCal dateCal due
Grade 0 gauge block set (steel, 1–100 mm)UKAS-0332-24-1187NPL
U=0.00021 mm (k=2, 95.45 %)
calibrated by UKAS 0332 · NPL
2026-06-012027-06-01

Specification provenance

Tolerance sourceMitutoyo 500-196-30 datasheet (AI-extracted, human-approved)
DocumentMitutoyo Digimatic Caliper 500 series datasheet, rev 2024-03, § Table 2 (accuracy ±0.02 mm ≤ 100 mm, ±0.03 mm ≤ 150 mm)
StatementTolerances derived from a manufacturer datasheet by AI extraction, reviewed and approved by a qualified person before use.

Electronic signatures (record hashes)

J. Moreno — Performed calibration
2026-09-09 04:47:57
sig:07cb7a7ecbbdcac1dfe784396a7db97ed9e2c40553b2b57af7371e4e98cf2d6f
J. Moreno — Approved calibration
2026-09-09 04:47:58
sig:20ed6596e812b19e8bb3080964a764cb8bb13a2f9da91a4576a8a778f39f9a45
The measurement results are metrologically traceable to the SI through the reference standards listed. Expanded uncertainties are reported at coverage factor k=2 (≈95 % confidence). Expanded uncertainties are shown to two significant figures; the test uncertainty ratio and any probability of non-conformance are computed from the unrounded uncertainty. Results relate only to the item calibrated. The calibration covers only the points listed; functions and ranges not listed were not calibrated. Where the item was not adjusted, the as-found and as-left values coincide and the stated uncertainty applies to both. Issued under the laboratory's accreditation UKAS 0417; the results reported are within the accredited scope unless marked otherwise.