<?xml version="1.0" encoding="UTF-8"?>
<dcc:digitalCalibrationCertificate xmlns:dcc="https://ptb.de/dcc" xmlns:si="https://ptb.de/si" schemaVersion="3.2.0">
  <dcc:administrativeData>
    <dcc:dccSoftware><dcc:software><dcc:name><dcc:content>Metroliq</dcc:content></dcc:name><dcc:release>1.0</dcc:release></dcc:software></dcc:dccSoftware>
    <dcc:coreData>
      <dcc:countryCodeISO3166_1>ZZ</dcc:countryCodeISO3166_1>
      <dcc:usedLangCodeISO639_1>en</dcc:usedLangCodeISO639_1>
      <dcc:mandatoryLangCodeISO639_1>en</dcc:mandatoryLangCodeISO639_1>
      <dcc:uniqueIdentifier>NCL-2026-00001/rev1</dcc:uniqueIdentifier>
      <dcc:identifications><dcc:identification><dcc:issuer>calibrationLaboratory</dcc:issuer><dcc:value>2026-09-09</dcc:value><dcc:name><dcc:content lang="en">Date of issue</dcc:content></dcc:name></dcc:identification></dcc:identifications>
      <dcc:beginPerformanceDate>2026-09-09</dcc:beginPerformanceDate>
      <dcc:endPerformanceDate>2026-09-09</dcc:endPerformanceDate>
      <dcc:performanceLocation>laboratory</dcc:performanceLocation>
    </dcc:coreData>
    <dcc:items><dcc:item>
      <dcc:name><dcc:content lang="en">Digital caliper 0–150 mm</dcc:content></dcc:name>
      <dcc:manufacturer><dcc:name><dcc:content>Mitutoyo</dcc:content></dcc:name></dcc:manufacturer>
      <dcc:model>500-196-30</dcc:model>
      <dcc:identifications>
        <dcc:identification><dcc:issuer>manufacturer</dcc:issuer><dcc:value>A2318804</dcc:value><dcc:name><dcc:content lang="en">Serial number</dcc:content></dcc:name></dcc:identification>
        <dcc:identification><dcc:issuer>customer</dcc:issuer><dcc:value>CAL-0091</dcc:value><dcc:name><dcc:content lang="en">Asset tag</dcc:content></dcc:name></dcc:identification>
      </dcc:identifications>
    </dcc:item></dcc:items>
    <dcc:calibrationLaboratory><dcc:contact><dcc:name><dcc:content>Northgate Calibration Ltd</dcc:content></dcc:name></dcc:contact></dcc:calibrationLaboratory>
    
    <dcc:respPersons>
        <dcc:respPerson><dcc:person><dcc:name><dcc:content>J. Moreno</dcc:content></dcc:name></dcc:person><dcc:role>Performed calibration</dcc:role></dcc:respPerson>
        <dcc:respPerson><dcc:person><dcc:name><dcc:content>J. Moreno</dcc:content></dcc:name></dcc:person><dcc:role>Approved calibration</dcc:role></dcc:respPerson></dcc:respPersons>
    <dcc:statements>
      <dcc:statement><dcc:name><dcc:content lang="en">Decision rule</dcc:content></dcc:name><dcc:declaration><dcc:content lang="en">guardband (guard bands per ILAC-G8 where applied)</dcc:content></dcc:declaration></dcc:statement>
      <dcc:statement><dcc:name><dcc:content lang="en">Accreditation</dcc:content></dcc:name><dcc:declaration><dcc:content lang="en">n/a</dcc:content></dcc:declaration></dcc:statement>
      <dcc:statement><dcc:name><dcc:content lang="en">Overall result</dcc:content></dcc:name><dcc:declaration><dcc:content lang="en">CONDITIONAL; worst-case TUR 1.73:1</dcc:content></dcc:declaration></dcc:statement>
    </dcc:statements>
  </dcc:administrativeData>
  <dcc:measurementResults><dcc:measurementResult>
    <dcc:name><dcc:content lang="en">Calibration results</dcc:content></dcc:name>
    <dcc:usedMethods>
          <dcc:usedMethod><dcc:name><dcc:content lang="en">Reference standard: Grade 0 gauge block set (steel, 1–100 mm) (cert UKAS-0332-24-1187), traceable to NPL</dcc:content></dcc:name></dcc:usedMethod>
          <dcc:usedMethod><dcc:name><dcc:content lang="en">Calibration procedure</dcc:content></dcc:name><dcc:description><dcc:content lang="en">NCL-DIM-004 · Calibration of digital calipers to 150 mm · rev 1</dcc:content></dcc:description></dcc:usedMethod></dcc:usedMethods>
    <dcc:influenceConditions><dcc:influenceCondition><dcc:name><dcc:content lang="en">humidity pct</dcc:content></dcc:name><dcc:data><dcc:text><dcc:content>44</dcc:content></dcc:text></dcc:data></dcc:influenceCondition><dcc:influenceCondition><dcc:name><dcc:content lang="en">temperature c</dcc:content></dcc:name><dcc:data><dcc:text><dcc:content>20.3</dcc:content></dcc:text></dcc:data></dcc:influenceCondition></dcc:influenceConditions>
    <dcc:comment>Condition as received: in_tolerance | Decision rule: Guard-banded acceptance (ILAC-G8:2019 §4.2.2 / ANSI-NCSL Z540.3 method 6): a result is reported PASS only when it lies inside the tolerance by at least the guard band w = U (U = expanded uncertainty, k as stated per point); a result within tolerance but inside the guard band is reported CONDITIONAL with its probability of false acceptance (PFA, JCGM 106); a result outside tolerance is reported FAIL. Probabilities assume a normal distribution with standard uncertainty u = U/k about the measured value (JCGM 106). Rule agreed with the customer before calibration. | Specification: Mitutoyo Digimatic Caliper 500 series datasheet, rev 2024-03, § Table 2 (accuracy ±0.02 mm ≤ 100 mm, ±0.03 mm ≤ 150 mm)</dcc:comment>
    <dcc:results>
          <dcc:result><dcc:name><dcc:content lang="en">10 mm</dcc:content></dcc:name>
            <dcc:data><dcc:list>
              <dcc:quantity><dcc:name><dcc:content lang="en">Nominal value</dcc:content></dcc:name><si:real><si:value>10</si:value><si:unit>\milli\metre</si:unit></si:real></dcc:quantity>
              <dcc:quantity><dcc:name><dcc:content lang="en">As-found</dcc:content></dcc:name><si:real><si:value>10.003</si:value><si:unit>\milli\metre</si:unit></si:real></dcc:quantity>
              <dcc:quantity><dcc:name><dcc:content lang="en">As-left (measured)</dcc:content></dcc:name><si:real><si:value>10.003</si:value><si:unit>\milli\metre</si:unit><si:measurementUncertaintyUnivariate><si:expandedMU><si:valueExpandedMU>0.01155</si:valueExpandedMU><si:coverageFactor>2</si:coverageFactor><si:coverageProbability>0.9545</si:coverageProbability></si:expandedMU></si:measurementUncertaintyUnivariate></si:real></dcc:quantity>
              <dcc:quantity><dcc:name><dcc:content lang="en">Conformity (as-left)</dcc:content></dcc:name><dcc:noQuantity><dcc:content>PASS</dcc:content></dcc:noQuantity></dcc:quantity>
              <dcc:quantity><dcc:name><dcc:content lang="en">Tolerance limits</dcc:content></dcc:name><dcc:noQuantity><dcc:content>9.98 … 10.02 mm (± 0.02 mm)</dcc:content></dcc:noQuantity></dcc:quantity>
              <dcc:quantity><dcc:name><dcc:content lang="en">Test uncertainty ratio</dcc:content></dcc:name><dcc:noQuantity><dcc:content>1.73:1</dcc:content></dcc:noQuantity></dcc:quantity>
              <dcc:quantity><dcc:name><dcc:content lang="en">Decision statement (as-left)</dcc:content></dcc:name><dcc:noQuantity><dcc:content>PASS (guard-banded)</dcc:content></dcc:noQuantity></dcc:quantity>
            </dcc:list></dcc:data></dcc:result>
          <dcc:result><dcc:name><dcc:content lang="en">50 mm</dcc:content></dcc:name>
            <dcc:data><dcc:list>
              <dcc:quantity><dcc:name><dcc:content lang="en">Nominal value</dcc:content></dcc:name><si:real><si:value>50</si:value><si:unit>\milli\metre</si:unit></si:real></dcc:quantity>
              <dcc:quantity><dcc:name><dcc:content lang="en">As-found</dcc:content></dcc:name><si:real><si:value>50.012</si:value><si:unit>\milli\metre</si:unit></si:real></dcc:quantity>
              <dcc:quantity><dcc:name><dcc:content lang="en">As-left (measured)</dcc:content></dcc:name><si:real><si:value>50.012</si:value><si:unit>\milli\metre</si:unit><si:measurementUncertaintyUnivariate><si:expandedMU><si:valueExpandedMU>0.01155</si:valueExpandedMU><si:coverageFactor>2</si:coverageFactor><si:coverageProbability>0.9545</si:coverageProbability></si:expandedMU></si:measurementUncertaintyUnivariate></si:real></dcc:quantity>
              <dcc:quantity><dcc:name><dcc:content lang="en">Conformity (as-left)</dcc:content></dcc:name><dcc:noQuantity><dcc:content>CONDITIONAL</dcc:content></dcc:noQuantity></dcc:quantity>
              <dcc:quantity><dcc:name><dcc:content lang="en">Tolerance limits</dcc:content></dcc:name><dcc:noQuantity><dcc:content>49.98 … 50.02 mm (± 0.02 mm)</dcc:content></dcc:noQuantity></dcc:quantity>
              <dcc:quantity><dcc:name><dcc:content lang="en">Test uncertainty ratio</dcc:content></dcc:name><dcc:noQuantity><dcc:content>1.73:1</dcc:content></dcc:noQuantity></dcc:quantity>
              <dcc:quantity><dcc:name><dcc:content lang="en">Decision statement (as-left)</dcc:content></dcc:name><dcc:noQuantity><dcc:content>CONDITIONAL — specification met (within tolerance) but inside the guard band, so NOT accepted under the guard-banded rule; probability of non-conformance if treated as accepted: 8.3% (normal, u = U/k)</dcc:content></dcc:noQuantity></dcc:quantity>
            </dcc:list></dcc:data></dcc:result>
          <dcc:result><dcc:name><dcc:content lang="en">100 mm</dcc:content></dcc:name>
            <dcc:data><dcc:list>
              <dcc:quantity><dcc:name><dcc:content lang="en">Nominal value</dcc:content></dcc:name><si:real><si:value>100</si:value><si:unit>\milli\metre</si:unit></si:real></dcc:quantity>
              <dcc:quantity><dcc:name><dcc:content lang="en">As-found</dcc:content></dcc:name><si:real><si:value>100.019</si:value><si:unit>\milli\metre</si:unit></si:real></dcc:quantity>
              <dcc:quantity><dcc:name><dcc:content lang="en">As-left (measured)</dcc:content></dcc:name><si:real><si:value>100.019</si:value><si:unit>\milli\metre</si:unit><si:measurementUncertaintyUnivariate><si:expandedMU><si:valueExpandedMU>0.01157</si:valueExpandedMU><si:coverageFactor>2</si:coverageFactor><si:coverageProbability>0.9545</si:coverageProbability></si:expandedMU></si:measurementUncertaintyUnivariate></si:real></dcc:quantity>
              <dcc:quantity><dcc:name><dcc:content lang="en">Conformity (as-left)</dcc:content></dcc:name><dcc:noQuantity><dcc:content>CONDITIONAL</dcc:content></dcc:noQuantity></dcc:quantity>
              <dcc:quantity><dcc:name><dcc:content lang="en">Tolerance limits</dcc:content></dcc:name><dcc:noQuantity><dcc:content>99.98 … 100.02 mm (± 0.02 mm)</dcc:content></dcc:noQuantity></dcc:quantity>
              <dcc:quantity><dcc:name><dcc:content lang="en">Test uncertainty ratio</dcc:content></dcc:name><dcc:noQuantity><dcc:content>1.73:1</dcc:content></dcc:noQuantity></dcc:quantity>
              <dcc:quantity><dcc:name><dcc:content lang="en">Decision statement (as-left)</dcc:content></dcc:name><dcc:noQuantity><dcc:content>CONDITIONAL — specification met (within tolerance) but inside the guard band, so NOT accepted under the guard-banded rule; probability of non-conformance if treated as accepted: 43.1% (normal, u = U/k)</dcc:content></dcc:noQuantity></dcc:quantity>
            </dcc:list></dcc:data></dcc:result>
          <dcc:result><dcc:name><dcc:content lang="en">150 mm</dcc:content></dcc:name>
            <dcc:data><dcc:list>
              <dcc:quantity><dcc:name><dcc:content lang="en">Nominal value</dcc:content></dcc:name><si:real><si:value>150</si:value><si:unit>\milli\metre</si:unit></si:real></dcc:quantity>
              <dcc:quantity><dcc:name><dcc:content lang="en">As-found</dcc:content></dcc:name><si:real><si:value>150.011</si:value><si:unit>\milli\metre</si:unit></si:real></dcc:quantity>
              <dcc:quantity><dcc:name><dcc:content lang="en">As-left (measured)</dcc:content></dcc:name><si:real><si:value>150.011</si:value><si:unit>\milli\metre</si:unit><si:measurementUncertaintyUnivariate><si:expandedMU><si:valueExpandedMU>0.01159</si:valueExpandedMU><si:coverageFactor>2</si:coverageFactor><si:coverageProbability>0.9545</si:coverageProbability></si:expandedMU></si:measurementUncertaintyUnivariate></si:real></dcc:quantity>
              <dcc:quantity><dcc:name><dcc:content lang="en">Conformity (as-left)</dcc:content></dcc:name><dcc:noQuantity><dcc:content>PASS</dcc:content></dcc:noQuantity></dcc:quantity>
              <dcc:quantity><dcc:name><dcc:content lang="en">Tolerance limits</dcc:content></dcc:name><dcc:noQuantity><dcc:content>149.97 … 150.03 mm (± 0.03 mm)</dcc:content></dcc:noQuantity></dcc:quantity>
              <dcc:quantity><dcc:name><dcc:content lang="en">Test uncertainty ratio</dcc:content></dcc:name><dcc:noQuantity><dcc:content>2.59:1</dcc:content></dcc:noQuantity></dcc:quantity>
              <dcc:quantity><dcc:name><dcc:content lang="en">Decision statement (as-left)</dcc:content></dcc:name><dcc:noQuantity><dcc:content>PASS (guard-banded)</dcc:content></dcc:noQuantity></dcc:quantity>
            </dcc:list></dcc:data></dcc:result>
    </dcc:results>
  </dcc:measurementResult></dcc:measurementResults>
</dcc:digitalCalibrationCertificate>